Sigma metrics analysis of serology screening assays to enhance quality and efficiency in New Zealand blood services

by | Jul 26, 2024 | Publications

Diagn Microbiol Infect Dis. 2024 Jul 17;110(2):116451. doi: 10.1016/j.diagmicrobio.2024.116451. Online ahead of print.

ABSTRACT

Sigma metric analysis was conducted across two New Zealand Blood Services (NZBS) laboratories (Auckland and Christchurch) to optimize quality control (QC) procedures. We evaluated five assays (anti-HCV, HIV Ag/Ab combo, HTLV-I/II, HBsAg, and Syphilis) using internal quality control (IQC) and third-party daily QC data extracted from four Architect i2000SR instruments during Jan 2 -31st, 2023. Mean, standard deviation (SD), and coefficient of variation (CV%) were calculated, assuming zero bias. Sigma metrics were determined using the Total Allowable Error (TEa %) based on difference between positive control mean and signal-to-cutoff (s/co) cut-off. Most assays exhibited CV% values ≤10 % except for HBsAg IQC (18.5 %) and anti-HCV third-party QC (13.4 %) at Christchurch. TEa % ranged from 38 % to 90 %. Overall, the assays demonstrated Six Sigma performance (σ > 6), except for HBsAg IQC (3.97) and anti-HCV third-party QC (5.46) at Christchurch. These high-quality serology assays can benefit from simplified QC design without compromising blood safety.

PMID:39053050 | DOI:10.1016/j.diagmicrobio.2024.116451